|
A
sampling system includes an aspiration line having an aspiration pump in series
with an aspiration probe. A pressure transducer 0-100 psi (26PCFFG6G) is used
after the sampling probe and connected to the pump via a pipette. When a test
sample is aspirated the valve is in the first valve position and the transducer
provides output voltage readings at predetermined time intervals during the
aspiration cycle that correspond to the vacuum level in the aspiration probe
during aspiration. A microprocessor is connected to the pressure transducer
compares the new/test aspirated sample output voltage to the one already stored
in the memory for unclotted sample serum. When the transducer output voltage
for the test sample differs from the reference voltage level by an amount
greater than two/three standard deviations of the reference voltage, it can be
determined that the test sample aspiration was impeded by a blood clot and
would not provide a reliable evaluation at the sampling system reading station.
The
discrepant test sample aspiration is thus unacceptable for test purposes and is
kept isolated from the analytical line by automatically changing the valve
position to prevent it from reaching the reading station and dispensing the
clotted sample to clean the sampling probe tip.
If
the difference between the test sample aspirations output voltage level and the
reference voltage level is less than two/three standard deviations of the
reference voltage level, the test sample is considered acceptable for test
purposes.
|